Ranging, 15us – YSI 3200 User Manual

Page 27

Advertising
background image

System Configuration and Operation

Section 3

Since the cursor is already in the Key beep field, press [UP] or [DOWN] to change the status of
the beeper. Press [MODE] two times to return to Operation mode.

RANGING

The Range menu allows the user to select between Auto and Manual ranging. The default is
autoranging, where the 3200 will automatically select the range that displays the highest
resolution for the sample being measured. Since this automatic selection takes time, the user may
select manual ranging when measuring several samples that fall into the same range.

From Operation mode, press the [SETUP] soft key, then press the [RANGE] soft key to enter the
Setup Range menu.

SETUP RANGE

[AUTO] Auto/Manual Range
[950 - 9999uS] Current range


UP

DOWN

NEXT

The Range menu allows the user to select between Auto and Manual ranging. Press the [UP] or
[DOWN] soft key to change from Auto to Manual. When ranging is set to Manual, the current
range can be set by pressing the [NEXT] soft key (to move the cursor to the Current range field),
then pressing the [UP] or [DOWN] soft key to select the desired range. The following hardware
ranges may be selected:

95.0mS - 3.00S

9.50 - 99.99mS

950 - 9999uS
95.0 - 999.9uS
9.50 - 99.99uS
0.950 - 9.999uS
0 - 0.9999uS

After setting up manual ranging, press the [MODE] key two times to return to Operation mode.

13.15uS




K=0.100
25.0°C

11:33

AM

24.38°

C

1.50%/°C
T CELL

CELL

SETUP

DIAGNOSTICS

RANGE

OPERATION

While making measurements (in Operation mode), the range can also be manually selected by
pressing the [RANGE] soft key. The current range selected will be displayed on the message line.