3 test and verification procedure – YSI 3200 User Manual

Page 64

Advertising
background image

Warranty and Repair

Section 9

9.3 TEST AND VERIFICATION PROCEDURE

Connect the YSI Model 3200 as shown in the Test and Verification Figure and follow the charts
below to verify its accuracy.

Equipment Required:

If using the YSI 3166 Precision Calibrator Set to simulate conductance, a YSI 3232 adapter is
required for proper connection to the cell connector. When using the 3232 adapter, temperature
can only be accessed via the EXT. TEMP connector which requires a .25” three conductor phone
plug.

If a decade resistance box is used to simulate conductance and temperature, a YSI #003229 cable
is required for connection to the cell connector.

Important Notes:

1. Temperature compensation must be turned off.

2. Put the 3200 in manual range mode.

3. Set the units to conductance and/or resistance.

4. The 3200 has no internal calibration. Opening the case should only be attempted by a

qualified service technician or permanent damage may result.

Resistance/Conductance Verification

Resistance Input

@ Cell Connector

3200 Range

Displayed Resistance

Displayed Conductance

1.000

Ω ± 0.5%

3.00 S *

1.000 ± .025

1.00 ± .04 S

10.00

Ω ± 0.2%

999.9 mS *

10.00 ± .119

100.0 ± 3.2 mS

100.00

Ω ± 0.1%

99.99 mS

100.0 ± 1.09

10.00 ± .11 mS

1000.0

Ω ± 0.1%

9999

µS

1.000 ± .011 K

1000 ± 11

µS

10.0 K

Ω ± 0.1%

999.9

µS

10.00 ± .11 K

100.0 1.1

µS

100.0 K

Ω ± 0.1%

99.99

µS

100.0 ± 2.1 K

10.00 ± .11

µS

1.000 M

Ω ± 0.1%

9.999

µS

1.00 ± .05 M

1.000 ± .021

µS

* No manual range change required. Instrument will display in correct range based on resistance
input

value.

YSI Incorporated

Model 3200

60