Appendix a - specifications – YSI 3200 User Manual

Page 71

Advertising
background image

APPENDIX A - SPECIFICATIONS

Modes

Conductivity

Conductance

Resistivity

Resistance

Salinity

TDS

Temperature

Conductance Range

Accuracy

Resolution

Frequency

0 - 0.9999

µS

± 0.30% full scale

0.0001

µS

40 Hz

0.950 - 9.999

µS

± 0.20% full scale

0.001

µS

80 Hz

9.50

µS - 99.99 µS

± 0.10% full scale

0.01

µS

290 Hz

95.0 - 999.9

µS

± 0.10% full scale

0.1

µS

1010 Hz

950 - 9999

µS

± 0.10% full scale

1

µS

1010 Hz

9.50 - 99.99 mS

± 0.10% full scale

0.01 mS

1010 Hz

95.0 - 999.9 mS

± 0.30% full scale

0.1 mS

1460 Hz

0.95 - 3.00 S

± 1.0% full scale

0.01 S

1460 Hz

Resistance Range

Accuracy Resolution

Frequency

0 - 9.999

± 0.2% full scale

0.001

1460 Hz

0 - 99.99

± 0.1% full scale

0.01

1460 Hz

0 - 999.9

± 0.1% full scale

0.1

1010 Hz

0 - 9.999 k

± 0.1% full scale

0.001 k

1010 Hz

0 - 99.99 k

± 0.1% full scale

0.01 k

1010 Hz

100.0 - 999.9 k

± 0.2% full scale

0.1 k

290 Hz

1.00 - 9.99 M

± 0.5% full scale

0.01 M

80 Hz

10.0 - 29.9 M

± 1% full scale

0.1 M

40 Hz

Salinity Range Accuracy

Resolution

0-80 ppt (NaCl)

±.1 ppt

0.1 ppt

TDS Range

Accuracy

Resolution

0-19,999 mg/L

0.50%

1 mg/L

Temperature Range

Accuracy Resolution

-5 - 100°C

±0.1°C

0.01°C

YSI Incorporated

Model 3200

67