Rear panel, Power supply connection, Cell connection – YSI 3200 User Manual

Page 9: Rs232 connection, External temperature connection

Advertising
background image

Introduction

Section

1

REAR PANEL

The rear panel contains the connections for the power supply, cell, external temperature probe
and RS232 serial port. It also has a mounting post for the 3232 adapter as shown below.

Adapter mounting post

POWER SUPPLY CONNECTION

The power supply connection requires a 12 VDC power supply (included) with at least 300 ma
current. The polarity is marked on the instrument.

CELL CONNECTION

The connector for the cell is a 7-pin mini DIN connector and is marked with an arrow to show
proper alignment. Be sure to align the arrows when plugging in the cell.

YSI 3200 series cells utilize a mini DIN connector that plugs directly into the 3200. If, however,
you have a YSI 3400 series cell, the YSI 3232 cell adapter and a YSI 3220 or YSI 700 series
temperature probe will be required. The YSI 3232 cell adapter mounts on the post on the rear of
the 3200. It has a 7-pin mini-DIN connector and two binding posts. The mini-DIN connector
plugs into the 3200 cell socket and a YSI 3400 series cell (or equivalent) connects to the binding
posts.

RS232 CONNECTION

The RS232 connection is a standard DB9 connector. This connector allows the 3200 to be
connected to the serial port of a computer or serial printer. A standard straight (NOT a null
modem) cable is required for connecting to a computer. To connect to a serial printer, use a
straight 9-pin male to 25 pin female cable (see Appendix - B).

EXTERNAL TEMPERATURE CONNECTION

The external temperature probe connection is used for temperature measurement and
compensation whenever the conductivity cell currently connected does NOT contain an internal