Autostable setup, 3 autostable setup, Mg /l – YSI 5000 User Manual

Page 19

Advertising
background image

16

5.3 Autostable Setup

The autostable feature indicates when readings are stable by emitting a single beep and
displaying an asterisk (*) to the right of the dissolved oxygen reading. The instrument uses
criteria that you input to determine what a stable reading is. You select the maximum percent of
change that may occur during a selected time duration. The default values are 10 seconds and
0.4% variation. The instrument will display an asterisk (*) only when these criteria are met.

8.20

*

mg
/L

24.38°C 11:33

AM

STORE

REVIEW

SEND

CALIBRATE

From the Setup menu, press the [STABLE] soft-key to display the following menu.

Stable

PARAMETERS

[ 10] Time Duration (3-19) Seconds

[0.4] DO % variation (0.0-1.9)

UP

DOWN

DIGIT

NEXT

This menu is used to change the autostable criteria. Press the [NEXT] soft-key to select between
the time duration and the percent variation. Use the [DIGIT] soft-key to select the digit you wish
to change by pressing [UP] or [DOWN]. When you have finished setting the parameters, press
[ENTER] to confirm. The STABLE sub-menu can be left without making any changes by
pressing the [MODE] key instead of [ENTER]. Operation returns to the setup menu.

If you wish to disable the autostable feature, set the DO% variation to 0.0.

To return to Main mode, press [MODE] two more times.

Main

Stable reading indication

Advertising
This manual is related to the following products: