Code/Problem Error

Possible

Cause

Correction

E4: Check

Membrane

DO High Current

Membrane too thin

Try another membrane

Electrodes

need

resurfacing

Repair or replace

Internal leakage in
probe or cable
connector

Repair or replace

Short circuit in probe or
cable

Repair or replace

E5: Check

Membrane

DO unstable

Membrane damaged

Replace membrane

Probe needs servicing

See probe instructions

Connector

improperly

installed

Check probe connection

Intermittent

connection

in cable or plug

Repair or replace

Probe malfunction

Repair or replace

ERROR:

DO reading < min. DO The DO reading at the start of the OUR or SOUR

measurement is below the “minimum beginning
DO” entered in the OUR or SOUR setup.
See 7.2 Application Mode, OUR, or SOUR.

ERROR:

Illegal value

The value entered is
outside the acceptable
range.

Enter the correct value

ERROR:

RTC defect control
register

Real time clock failure

Return for repair

ERROR:

Init. Logger failed

Logger initialization
failed

Return for repair