YSI 5000 User Manual

Page 37

Advertising
background image

34

2.6

mg
/L /h

23.21°C 8.92

0:00

OUR SETTING SAVED

START

SETUP

REVIEW

SEND

Once the parameters have been set, you are ready to begin. Place the probe in the prepared
sample and make certain no air bubbles are trapped, then turn on the stirring and wait a few
seconds for the temperature readings to stabilize.

NOTE: It is normal for the OUR reading to oscillate for the first few seconds.

Press [START] to begin the OUR measurement. The following screen shows the available soft-
keys after the test is started.

348.7

mg
/L /h

23.20°C 6.97

0:32

END

SEND ON

The OUR value is displayed on the screen and is updated approximately once per second.

The [SEND ON/OFF] soft-key toggles the RS232 status. Press [SEND ON] if you want to send
the readings to a computer or serial printer during the OUR measurement. The RS232 status is
displayed just above the elapsed time. During the OUR measurement, data will be sent every 15
seconds until the OUR measurement ends or the [SEND OFF] soft-key is pressed. The following
is an example of the format:

NOTE: Once pressed, the [SEND ON] soft-key toggles and becomes a [SEND OFF] soft-key.

Second mg/L/h

C

mg/L 11:04:10 09/21/98

0

0.23 23.20

8.52

15 595.27 23.20

7.28

30 369.63 23.20

6.98

45 275.23 23.20

6.80

. . .

mg

/L

OUR

Current DO

OUR

OUR

mg

/L

Elapsed Time

SEND: off

RS232 status

RS232 status

SEND: off

Elapsed Time