YSI 5000 User Manual

Page 41

Advertising
background image

38

Press [SETUP] to change the SOUR parameters. The following screen shows the default
parameters.

Sample / Total 1 / [ 1]

SOUR

Min. Time

[ 1] min.

Max. Time

[ 15] min.

[*]SOUR@20

Min. Beginning DO

[ 5.00] mg/L SEND:

off

Min. Ending DO

[ 2.00] mg/L

Solids Weight

[ 1.000] g/L

UP

DOWN

DIGIT

NEXT

Use the [UP], [DOWN], [DIGIT] and [NEXT] soft-keys to change the following parameters as
necessary:

Sample / Total:

Enter the ratio of sample volume to total volume. For example, if
you are diluting 1 to 10 (1 part sample plus 9 parts dilution water)
then enter the number 10. This would be a 10-fold dilution.

Min. Time:

Enter the minimum time (in minutes).

Max. Time:

Enter the maximum time (in minutes). When the maximum time is
reached, the test will end.

Min. Beginning DO: Enter the minimum level of DO allowed at the start of the test. If the

DO falls below this level, the test will not start.

Min. Ending DO:

Enter the minimum level of DO allowed during the test. If the DO
level falls to this value, the test will end.

Solids Weight:

Enter the Total Solids or Volatile Suspended Solids concentration of
the sample in g/L. The maximum acceptable value is 31.999 g/L.

SOUR@20:

This applies the calculation for temperature correction to 20°C
according to the Farrell and Bhide equation as follows:

SOUR

20

= SOUR

T

x Θ

(20-T)

Where:

Θ = 1.05 above 20°C

1.07 below 20°C

This calculation is only valid for temperature ranges from 10° - 30°C

SEND: on/off

This sets the RS232 status at the beginning of the test.

After the parameters have been set, press [ENTER] to save them and return to the SOUR menu.