From the Model 5100 Application mode menu press the [OUR] (Oxygen Uptake Rate) soft-key.
The following screen will be displayed.

4.7

mg
/L /h

23.21°C 8.92

0:00

START

SETUP

REVIEW

SEND

Press [SETUP] to change the OUR parameters. The following screen shows the default
parameters.

PARAMETERS

OUR

Sample / Total 1 / [ 1]

Min. Time

[

1] min.

Max. Time

[

15] min.

SEND: off

Min. Beginning DO

[ 5.00] mg/L

Min. Ending DO

[ 2.00] mg/L

UP

DOWN

DIGIT

NEXT

Use the [UP], [DOWN], [DIGIT] and [NEXT] soft-keys to change the following parameters as
necessary:

Sample / Total:

Enter the ratio of sample volume to total volume. For example, if
you are diluting 1 to 10 (1 part sample plus 9 parts dilution water),
then enter the number 10. This would be a 10-fold dilution.

Min. Time:

Enter the minimum time (in minutes).

Max. Time:

Enter the maximum time (in minutes). When the maximum time is
reached, the test will end.

Min. Beginning DO: Enter the minimum level of DO allowed at the start of the test.

Min. Ending DO:

Enter the minimum level of DO allowed during the test. If the DO
level falls below this value, the test will end.

SEND: on/off

This sets the RS232 status at the beginning of the test.

After the parameters have been set, press [ENTER] to save them and return to the OUR menu.

OUR

mg

/L

SEND: off