54

12. Troubleshooting

Error Messages
The instrument performs a Power On Self Test each time it is turned on. The following error
codes are provided to facilitate troubleshooting.

Code/Problem Error

Possible

Cause

Correction

Blank Display

Dead/low batteries

Replace batteries or
connect power supply

5011

adapter

improperly connected

Unplug 5011 from
power supply connector

E1: Check Probe

Temperature under
range

Connector improperly
installed

Check probe
connection

Intermittent

connection

in cable or plug

Repair or replace

Temperature

over

range

Electrical leakage in
connector or cable

Repair or replace

Faulty

temperature

sensor

Repair or replace

E2: Check

Membrane

High Background
current (DO)

Insufficient warm-up
time

See Calibration

Incorrect probe zeroing
procedure

See Zero Calibration

Probe needs servicing

See probe instructions

Probe malfunction

Repair or replace

E3: Check

Membrane

DO Low current

Insufficient electrolyte

See probe instructions

Contaminated

electrodes or fouled
membrane

See probe instructions

Membrane too thick

Try another membrane

High resistance in
probe connection

Repair or replace

Advertising