REVIEW

20.1 mg /h /g @20

6.8 mg /L

REC:1 EMPTY

10:32 AM

09/21/98

NEXT REC.

MODIFY

SAVE

SEND

Press the [NEXT REC.] soft-key to change the record number to find an empty slot, or pick a slot
to overwrite (you cannot delete results, you must overwrite). Press the [SAVE] soft-key to save
the current test data into that slot. Once you have saved the data you can press the [MODIFY]
soft-key to change the parameters of the test.

MODIFY

20.1 mg /h /g @20

6.8 mg /L

REC:1 FULL

10:32 AM

09/21/98

!

"

SETTINGS

Press the ! or " soft-keys to move the start time reference point to an optimal place on the
curve. The reference point is represented by a vertical line. Only data to the right of the
reference point will be used to calculate the SOUR value. Moving this reference point allows
you to start the test at anytime and trim off unwanted data after the test has run. The DO reading
at the reference point is shown at the top. The reference point cannot move closer than the
minimum time to the end of the test.

The [SETTINGS] soft-key allows you to modify or input values for the Dilution factor and
Solids Weight. These changes are saved if you save the record again.


0

mg
/L

705

SOUR

DO (at Reference

Test time

Time Reference

0

mg
/L

705

Record Number

FULL or EMPTY
indicator