Auto Cal
1. Prepare the probe according to the instructions in the Probe Operations Manual.

2. Connect the probe to the meter.

3. Place the probe in a BOD bottle containing about 1

of water to provide a 100% relative

humidity environment.

4. Press [ ] to turn the instrument on.

5. Allow the probe to polarize and the temperature to stabilize for at least 15 minutes. If

calibration is performed prematurely the values will drift and may be out of specification.

6. Press the [CALIBRATE] soft-key to change to Calibration mode. The following display will

appear.

98.0

*

% 8.15

mg/L

Calibrate

0.0

ppt

740

mmHg

24.38°C 11:33

AM

AUTO CAL

DO CAL

SETUP

DIAGNOSIS

7. Model 5100: Verify that the barometer reading is correct. Calibrate the barometer, if

necessary, as shown under 6. Dissolved Oxygen Calibration, Barometer.

Model 5000: Enter the current barometric pressure as shown under 6. Dissolved Oxygen

Calibration, Barometer.

8. Make sure that the display readings are stable, then press the [AUTO CAL] soft-key to

calibrate Dissolved Oxygen. The message “

D.O. CALIBRATION SAVED

” will be displayed for

a few seconds.

98.0

*

% 8.15

mg/L

Calibrate

0.0

ppt

740

mmHg

24.38°C 11:33

AM

D. O. CALIBRATION SAVED

AUTO CAL

DO CAL

SETUP

DIAGNOSIS

9. Press [MODE] to return to the Main mode. The instrument is now calibrated and ready to

measure dissolved oxygen and temperature. See 7.1 Main Mode, Making Measurements.