Do cal (manual) – YSI 5000 User Manual

Page 24

Advertising
background image

21

DO Cal (manual)
1. Prepare the probe according to the probe instructions.

2. Connect the probe to the meter.

3. Place the probe in a known oxygen environment, such as a BOD bottle containing about 1

of

water to provide a 100% relative humidity environment or a Winkler-titrated sample.

4. Press [ ] to turn the instrument on.

5. Allow the probe to polarize and the temperature to stabilize for at least 15 minutes. If

calibration is performed prematurely the values will drift and may be out of specification.

6. Press the [CALIBRATE] soft-key to change to Calibration mode. The following screen will

be displayed.

98.0

*

% 8.15

mg/L

Calibrate

0.0

ppt

740

mmHg

24.38°C 11:33

AM

AUTO CAL

DO CAL

SETUP

DIAGNOSTICS

7. Press the [DO CAL] soft-key to enter the manual DO calibration menu.

98.0

*

% 8.15

mg/L

Cal %

0.0

ppt

740

mmHg

24.38°C 11:33

AM

UP

DOWN

DIGIT

NEXT

8. Make sure that the display readings are stable, then enter the calibration value in percent

saturation, using the [UP], [DOWN] and [DIGIT] soft-keys. See Appendix A - Oxygen
Solubility Table
and
Appendix B - Pressures and Altitudes for manual calibration data.

If you wish to calibrate in milligrams per liter (instead of percent), verify that the salinity is set to

the salinity value of the calibration environment (0.0 ppt for air calibration, see 6.1 Dissolved
Oxygen Calibration, Salinity
). Then, use the [NEXT] soft-key to select mg/L and enter the
calibration value in milligrams per liter as shown below.

Flashing digit

Advertising