History – YSI 5000 User Manual

Page 48

Advertising
background image

45

12. Troubleshooting for additional information on error codes.

From the Main mode, press the [CALIBRATE] soft-key to enter Calibration mode, then press the
[Diagnosis] soft-key to enter Diagnosis mode. The following screen will be displayed.

Diagnosis

HISTORY

SENSOR

HELP

RESET Ps

History
Press the [HISTORY] soft-key to display the following information.

Date Time DO

uA

%/uA C Diagnosis

02/01 14:12 16.8 5.98 21.9

02/02 08:42 16.3 7.90 28.0

02/05 09:55 17.0 6.42 23.0

02/06 10:58 17.1 7.47 27.9

02/07 13:08 16.9 6.43 22.8

HISTORY

SENSOR

HELP

RESET Ps

The History screen displays information about the last five DO calibrations. This information is
useful for tracking the performance of the DO probe/membrane. The following parameters are
displayed:

Date:

The date the calibration was performed.

Time:

The time the calibration was performed.

DO uA:

The dissolved oxygen calibration current in microamps. The calibration
current (100% saturation at sea level) of an ideal probe using a 1 mil
(standard) membrane is 13.7uA @ 20°C. If the probe current (100%
saturation @ 20°C, 1 mil membrane) is greater than 17.0uA or lower than
8.0uA, the membrane/probe may need service. See 12. Troubleshooting,
Error Codes, E3 and E4.

%/uA:

The slope of the dissolved oxygen probe in percent per microamp
corrected to 20°C. The slope of an ideal probe using a 1 mil membrane is