Sensor – YSI 5000 User Manual

Page 49

Advertising
background image

46

7.3 %/uA @ 20°C. If the probe slope (1 mil membrane) is greater than
12.6 %/uA or lower than 5.9 %/uA, the membrane/probe may need
service. See 12. Troubleshooting, Error Codes, E3 and E4.

C:

The temperature at the time of calibration in °C.

Sensor
Press the [SENSOR] soft-key to display the following sensor diagnostics screen:

System Parameters

Sensor

Cal. Temperature

=

22.85 C

Probe Slope

=

6.43 %/uA @20C

Probe Offset

=

0.000 uA

Probe Current

=

16.91 uA

Pressure Offset

=

830 mbar

P = 981

Cal. Temperature:

The temperature at the time of the last calibration in °C.

Probe Slope:

The slope of the dissolved oxygen probe in percent per microamp
corrected to 20°C. The slope of an ideal probe using a 1 mil
membrane is 7.3%/uA @ 20°C. If the probe slope (1 mil
membrane) is greater than 12.6 %/uA or lower than 5.9 %/uA, the
membrane/probe may need service. See 12. Troubleshooting, Error
Codes, E3 and E4.

Probe Offset:

The zero offset current (background current) of the DO probe in
microamps. The factory default value is 0.000 uA. This value will
only change after a zero calibration is performed (see 6.2 Zero
Calibration
). I
f this value is greater than 0.15uA (after a zero
calibration) the membrane/probe may need service. See 12.
Troubleshooting, Error Code E2
.

Probe Current:

The present (real time) DO probe current in microamps.

Pressure Offset:

The offset of the barometer in millibars. This value will change
whenever the barometer is calibrated. The normal range is 300 to
1300 mbar.

P:

Current barometer reading in millibars.

Press [MODE] to return to the Diagnosis menu. Press [MODE] two more times to return to the
Main menu.