Appendix d - remote command language, Command format, Prefix – YSI 5000 User Manual

Page 68: Suffix, Command types and parameters

Advertising
background image

65

19. Appendix D - Remote Command Language

Remote commands and data output are available only when the instrument is in the Remote
mode.

Command Format:
<Prefix><Command Type><Parameter><Suffix>

Prefix:
The prefix consists of the two character sequence <ESC>&. In HEX this is 1B 26.

Suffix:
The suffix consists of the two character sequence <CR><LF>. In HEX this is 0D 0A.

Command Types and Parameters:
Command types are arranged into two groups, immediate and automatic.


The Display command controls the display of remote messages.

Command type parameter description

D

0

string

echoes the string to the first line of the display

D

1

string

echoes the string to the second line of the display

D

2

none

Clear display lines 0 and 1

D

3

0

Set display update flag line 0

D

3

1

Set display update flag line 1


The ouTput mode command controls how data flow is controlled. A bar code scan in T1 or T2
mode will cause a line of output, just as if the T command had been received.

Command type parameter description

T

0

none

Continuous data output

T

1

none

One line of data output in response

T

2

none

One line after stable reading

Advertising