Appendix f advanced calibration setup – YSI 6500 Environmental Process Monitor User Manual

Page 106

Advertising
background image

APPENDIX F

ADVANCED CALIBRATION SETUP


The Calibration Setup submenu of the 6500 Main menu presents two options, both of which are
covered in Setup, Section 3.4. The basic function of each is outlined below.

(1) Calibrations enabled is a list of all possible sensor calibrations which can either be enabled
or disabled with regard to calibration from the Run mode.

(2) Advanced setup is a selection which allows the user to choose whether calibration errors can
be overridden and defines the default calibration value which will appear on the display during
the calibration protocol.

The following description provides additional information related to using the Advanced setup
option. From Main menu highlight Calibration setup and press Enter, then Advanced setup
and press Enter as shown in the screens below.










After entering the Advanced setup submenu a display similar to the following for specific
conductance will appear.









You need to consider four options within this menu.

Is: Enabled. This determines whether or not the calibration of this parameter is enabled (allowed)
from the Run mode of the 6500 display. Highlight the selection “Is: Enabled”. Press Enter to
toggle back and forth between “Is: Enable” and “Is: Disabled”. Select the desired option.