Accessories Appendix

D

6510 – PANEL MOUNT KIT


Panel mounting the 6500 Monitor is a simple process using the enclosed mounting hardware and
the following tools; 5/32” Allen wrench, Philips screwdriver, and the necessary tools for cutting
the mounting hole in the control panel. The 6500 Monitor may be mounted in any panel with
9.5”L x 7.5”W space available, and behind the panel depth of 5.5”.

The location of the 6500 Monitor should be elevated and in a dry place above potential flood
level. The Monitor should be easily accessible for an operator or technician to have convenient
access.

The 6500 should not be mounted on hot or vibrating pipe or structure, near a high heat source,
near an AC motor or transformer, radio transmitter or antenna.

The following steps should be followed when panel mounting the 6500 Monitor.

1. Cut the mounting hole in the panel where the 6500 will be mounted using dimensions from

Figure 21.

2. Remove the 6500 lid by loosening the four corner screws, and remove the conduit fittings

from the bottom of the 6500 during installation.

3. Securely fasten both mounting brackets (included the ones included with the 6500) to the

back of the 6500 enclosure with the countersink screws provided as shown in Figure 22.

4. Attach the mounting flange using the, 4” standoffs, Nylon spacers, Allen head screws, and

lock-washers as shown in Figure 23.

Note: Insure the Nylon spacers are positioned with the small end inserted into the front of
the brackets to correctly position the standoffs and prevent the mounting flange from
warping, as indicated in Figure 23.

5. Attach the mounting plate to the panel using the sheet metal screws in the four corner holes.

6.

Ensure that all hardware is tightened.

YSI Incorporated

6500 Environmental Process Monitor

91