3 installing the sonde – YSI 6500 Environmental Process Monitor User Manual

Page 11

Advertising
background image

Installation Section

2

Figure 3 Remote location of monitor relative to sonde

Enter

Esc

Cal

6500

6500

EN VIRON MEN TAL

EN VIRON MEN TAL

MO N ITORIN G

MO N ITORIN G

S YS TEM

S YS TEM

21.2 Temp
7.35 DO
6.53 pH

sonde cable

w/ MS-8

Stream

Sonde

Breakout Box

Conduit &

6507

6’ Patch cable

w/ MS-8

customer-supplied
3-conductor cable
250’ Maximum

Junction Box

6508

6500 Monitor

Not drawn to scale.

AC


YSI also provides a means for convenient and accurate bench calibration of the sonde. Refer to
the Sonde Manual for bench calibration procedures. Bench calibration allows the operator to
easily disconnect the sonde from the monitor and then reconnect the sonde to a laboratory
computer or terminal device (YSI 610 display/logger). Under this protocol, the sonde can be
fully checked in a user-friendly environment, recalibrated and then redeployed. If the sonde is
equipped with a bulkhead connector rather than an integral cable, bench calibration is used.

If locating the monitor near the sonde is not possible due to environmental conditions and/or
accessibility, bench calibration provides a very good alternative to on-line calibration.

Although the monitor is suitable for outdoor use, it should be located in an area where
temperature extremes, vibrations, electromagnetic and radio frequency interference are minimal.
Select an installation location that is at least two (2) feet from any high voltage conduit. Avoid
mounting on severely vibrating structures. Be sure the monitor can be fully opened and serviced
at its installed location by maintenance personnel. Be sure the site has easy access for operating
personnel.

2.3 INSTALLING THE SONDE


After you have chosen suitable sites for the monitor and sonde, proceed with the installation as
described below.

Avoid routing sonde cabling near wiring associated with rotating machinery and/or equipment
involving electrical switching or regulation. Consider placing sonde cabling in grounded metallic
conduit if unstable readings appear due to electromagnetic interference.

YSI Incorporated

6500 Environmental Process Monitor

7