6504, front cover removed, 2) black common (3) purple sdi-12 – YSI 6500 Environmental Process Monitor User Manual

Page 87

Advertising
background image

Accessories Appendix

D

Enter

Esc

Cal

6500

6500

EN VIRON MEN TAL

EN VIRON MEN TAL

MON ITORIN G

MON ITORIN G

SYSTEM

SYSTEM

21.2 Temp
7.35 DO
6.53 pH

6507

6’ Patch cable

w/ MS-8

Sonde cables

w/ MS-8

Stream

Sondes

6504

Breakout Box

6507

Gland Fitting

Enter

Esc

Cal

6500

6500

EN VIRO N MEN TAL

EN VIRO N MEN TAL

MO N ITO RIN G

MO N ITO RIN G

SYSTE M

SYSTE M

21.2 Temp
7.35 DO
6.53 pH

Junction Box

6508

Sonde cables

w/ MS-8

Stream

Sondes

Breakout Box

Conduit &

6507

6’ Patch cable

w/ MS-8

customer-supplied
3-conductor cable


In remote Breakout Box installations, Figure 4, the 6500 Monitor uses a standard MS-8
connection to interface with the 6 foot (1.8 m) 6507 Patch Cable that is connected in the 6508
Junction Box. The Junction Box is then connected by conduit and three- conductor cable to the
Breakout Box(s). The customer is advised to supply rigid conduit and 18 AWG or heavier
shielded multi-conductor cable to connect between the Junction Box and the Breakout Box at the
Sonde installation site. This configuration allows a maximum cable length of 250’ between the
6500 Monitor and the Sonde(s). The cable from the junction box should be landed to TB-1 in the
breakout box. The MS-8 connectors for Sonde hookups are pre-wired to TB-2 and TB-3. See
Figure 12 for further information.


6504, Front Cover Removed

(2) MS-8 to Sonde

3/4” conduit to Jct. Box


The TB-1, Figure 12, conductor color
code is:
(1) Red

+12 VDC

(2) Black

Common

(3) Blue

SDI-12

IMPORTANT!

GND on TB-1 is for signal common only.

Do not connect to earth ground.

1

3

2

TB-1

Figure

11

Figure 10


The TB-2 & TB-3, Figure 12, conductor color code is:


(1) Red

+12 VDC

Figure12

(2) Black

Common

(3) Purple

SDI-12


YSI Incorporated

6500 Environmental Process Monitor