4.8.2

Behavior in non-operative condition

An output is non-operative when no function is activated for the output.

An output becomes non-operative in case of

Power failure
(As soon as the supply voltage is sufficient again, the non-operative
condition of the outputs ends. The outputs function as specified by
the user again.)

Erasing a link to a sensor

Changing the Measuring mode sensor setting for a linked sensor

Changing the Measuring range sensor setting for a linked sensor

Note
Before editing sensor settings a note appears on the display to inform
you that links will be erased when you change the Measuring mode or
Measuring range sensor setting.

Settings in the non-

active condition

Current output

Relay output

Current: 0 A

Relay: Open