YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 125

Advertising
background image

System 182

Outputs

6 - 29

ba76022e01

12/2011

Application example

 Feedback control of the D. O. concentration

 Sensor: TriOxmatic 700 IQ (measuring range: 0 ... 60 mg/l)

The control parameters form the following (negative) characteristic
curve:

Fig. 6-14 Characteristic curve for application example

The controller works with the following gain:

Control parameters

Value

Nominal value

4 mg/l

Xp

10 % of the measuring range or 6 mg/l

Imin

8 mA

Imax

14 mA

Io

12 mA

ti

0 s (no I controller part)

td

0 s (no D controller part)

0

8

14

12

Imax

Imin

Current I [mA]

Measured value [mg/l]

4

2

8

Io

Nominal value

Xp

mg/l

mA

1

mg/l

6

mA

6

K

=

=