Electrical data iq s, 24v (24 v ad/dc version), Electrical connections iq s – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 139: Ensor

Advertising
background image

System 182

Technical data

9 - 3

ba76022e01

12/2011

Electrical data

IQ S

ENSOR

N

ET

/24V

(24 V AD/DC version)

Electrical connections

IQ S

ENSOR

N

ET

(line power version)

Terminal strip inside the enclosure:

Electrical connections

IQ S

ENSOR

N

ET

/24V

(24 V AC/DC version)

Terminal strip inside the enclosure:

Supply

Nominal voltage:

24 V AC/DC ± 10 %
protective low voltage
SELV (Safety
Extra Low Voltage)

AC frequency:

50/60 Hz according to
DIN IEC 60038

Connection:

2 pin

Line cross-section of connections:

Europe:

1.5 ... 4.0 mm

2

USA:

AWG 14 ... 12

Fuse rating on the operator side: max. 16 A

Power consumption

max. approx. 12 W

SENSORNET 1

RED

SHIELD

GREEN

X3

X2

X17

X9

X15

X13

X7

X11

X5

X1

X16

X8

X14

X12

X6

X10

X4

100...

240V AC

MAINS

AUXILIARY

R3

R2

R1

L1

N

240V AC

2A AC

240V AC

2A AC

240V AC

2A AC

AUXILIARY

VOLTAGE

0/4...20mA

+ REC -

C2

0/4...20mA

+ REC -

C1

SENSORNET 1

RED

SHIELD

GREEN

X3

X2

X17

X9

X1

X16

X8

24V AC DC

INPUT

POWER

AUXILIARY

AUXILIARY

VOLTAGE

X7

X5

X6

X4

0/4...20mA

+ REC -

C2

0/4...20mA

+ REC -

C1

X15

X13

X11

X14

X12

X10

R3

R2

R1

240V AC

2A AC

240V AC

2A AC

240V AC

2A AC

Advertising