System 182
Installation
3 - 31
ba76022e01
12/2011
Preparing the power
cable
Fig. 3-25 Prepared power cable.
Caution
The ground wire must not project into the enclosure. Otherwise,
malfunctions could occur.
1
Cut off the cable to the required length.
2
Strip the cable insulation for approx. 45 mm.
3
Bare the wires of phases L and N and fit them with wire end
sleeves.
4
If present, cut off the ground wire at the end of the cable
sheath.
Advertising