YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 140
Technical data
System 182
9 - 4
ba76022e01
12/2011
Relay
(3 x)
Current outputs
(2 x)
Output
Galvanically separated
Max. switching voltage
240 VAC or 24 VDC
Max. switching current
2 A (AC and DC)
Installation
requirements
Fuse rating on the operator side: max. 2 A
Relay functions
Programmable as:
– Opener or closer
– Limit monitor
– Monitoring of the warning and error
signals of the IQ S
ENSOR
N
ET
– Proportional frequency output
– Proportional pulse width output
Output
Galvanically separated from the sensors
Output current
Can be switched between 0 - 20 mA and
4 - 20 mA
In the case of errors, can be set to: 0 ... 21
mA
Max. initial output
voltage
24 V, in the case of missing or incorrect
burden
Accuracy
0.3 % of current value ± 50 µA,
load max. 500
Ω
Functions
Programmable:
– Recorder attenuation, adjustable
0 - 40 mA/s
– Error behavior can be set as required
0 ... 21 mA
– Error behavior according to Namur NE43
can be selected
– Positive and negative characteristic
curve
– PID controller