1 switching on the maintenance condition, 2 switching off the maintenance condition, Switching on the maintenance condition -42 – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 214: Switching off the maintenance condition -42, Ee section 4.9.2, Section 4.9.1), See section 4.9.2

Advertising
background image

Settings

MIQ/CR3; DIQ/CR3

4 - 42

ba76032e01

01/2012

4.9.1

Switching on the maintenance condition

4.9.2

Switching off the maintenance condition

1

Call up the measured value display with

m

.

2

Select the sensor you want to switch on the maintenance
condition for with

d

. The measured values of the sensor do

not flash.

3

Open the 300 - Display/Options menu with

g

.

4

Using

d

, highlight the menu item, Switch maintenance

condition on/off or Maintenance of sensor S0x and confirm with

g

. A window that informs you about the maintenance

condition opens up.

5

Confirm Continue with

g

.

The selected sensor is in the maintenance condition. Linked
outputs are frozen.

6

Call up the measured value display with

m

.

The measured values of the sensor flash.

1

Call up the measured value display with

m

.

2

Select the sensor you want to switch off the maintenance
condition for with

d

. The measured values of the sensor

flash.

3

Open the 300 - Display/Options menu with

g

.

4

Using

d

, highlight the menu item, Switch maintenance

condition on/off or Maintenance of sensor S0x and confirm with

g

. A window that informs you about the maintenance

condition opens up.

5

Confirm Continue with

g

.

The maintenance condition of the selected sensor is switched
off. Linked outputs are released.

6

Call up the measured value display with

m

.

The measured values of the sensor do not flash.