8 indexes, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 221: 2 info messages, Indexes -1, Explanation of the messages -1 8.1.1, Error messages -1, Info messages -1, 8indexes

Advertising
background image

MIQ/CR3; DIQ/CR3

Indexes

8 - 1

ba76032e01

01/2012

8

Indexes

8.1

Explanation of the messages

In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S

ENSOR

N

ET

system for the MIQ/CR3 output module.

Note
Information about

Contents and structure of the log book and

Structure of the message code

can be found in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system

operating manual.

Note
All message codes of the output module MIQ/CR3 or DIQ/CR3 end
with the number "411".

8.1.1

Error messages

8.1.2

Info messages

The MIQ/CR3 or DIQ/CR3 output module does not send any info
messages.

Message code

Message text

EA4411

Current output range undercut
* Check process
* Check settings and, if necessary, change them

EA5411

Current output range exceeded
* Check process
* Check settings and, if necessary, change them

EI3411

Burden resistor too large (> 500 Ohm) or current loop interrupted
* Check burden, terminal connections and connection lines

Advertising