Using DIQ modules (accessories)

Note
The various application possibilities of the DIQ modules are shown by
means of examples in section 3.11.

3.6.1

DIQ/JB

The DIQ/JB module is a passive branching module and can be used for
the following purposes

 To extend the SACIQ sensor connection cable, e.g. to connect a

sensor that is located farther away to the Universal Transmitter.

 To branch a line at the end of an extension.

Fig. 3-22 DIQ/JB open.

The DIQ/JB module has seven potential free terminals. To extend or
branch lines, connect the three IQ S

ENSOR

N

ET

wires to each other

one-to-one at any terminals

 green <-> green

 red <-> red

 black/filler stranded wire <-> black/filler stranded wire.

3.6.2

DIQ/CHV

The DIQ/CHV module is a valve module for the automatic relay-
controlled compressed air-driven cleaning function in the 182 system.
It provides four additional potential free terminals to branch (extend)
interface lines. For each sensor that is to have compressed air cleaning
a DIQ/CHV is required.