YSI DIQ/S 182-MOD System 182 User Manual

Page 125

Advertising
background image

System 182

Modbus connection

7 - 13

ba76025e01

02/2012

12

42h

Contents of register 13
(HI)
= main measured
value

429E46C2h (Float 32)
--> 79,1382
Measured parameter and
unit,
see byte 10 (measuring
mode)

13

9Eh

Contents of register 13
(LO)
= main measured
value

14

46h

Contents of register 14
(HI)
= main measured
value

15

C2h

Contents of register 14
(LO)
= main measured
value

16

00h

Contents of register 15
(HI)
= secondary measured
value

00000000h (Float 32)
--> 0
but measured value
invalid (MISSING),
see Byte 11 (measured
value status)

17

00h

Contents of register 15
(LO)
= secondary measured
value

18

00h

Contents of register 16
(HI)
= secondary measured
value

19

00h

Contents of register 16
(LO)
= secondary measured
value

20

23h

CRC (HI)

Checksum (CRC)

21

5Eh

CRC (LO)

Modbus response

Byte

Value

Information

Meaning of the value