YSI DIQ/S 182-MOD System 182 User Manual

Page 27

Advertising
background image

System 182

Installation

3 - 13

ba76025e01

02/2012

Preparing the cable

ends

Fig. 3-11 Prepared cable end

Connecting the cables

The SNCIQ and SNCIQ/UG cables are connected to the terminal strip
in the same way as the SACIQ sensor connection cable (see
section 3.4.5):

1

Cut off the cable to the required length.

2

Remove approx. 45 mm of cable insulation (in the case of the
SNCIQ/UG earth cable, remove both the inner and outer
insulation).

3

Only for the SNCIQ/UG earth cable:
strip the outer insulation for a further 35 mm.

4

Shorten the exposed shielding braid up to the cable sheath.

5

Shorten the two fillers (plastic inlays) up to the cable sheath.

6

Bare the red and green wires and fit them with wire end
sleeves.

7

Fit the filler stranded wire with a wire end sleeve.

SNCIQ

SNCIQ/UG

approx. 35 mm

approx. 45 mm

approx. 45 mm

1

Open the enclosure of the Universal Transmitter or
MIQ module.

2

Select a free SENSORNET connection. At the same time, look
out for the SENSORNET designation on the label on the
bottom of the enclosure.