4.6
Passwords
You can assign and activate two passwords for the System 182.
Settings password
protects all settings in the SETTINGS menu.
If the password is active, all settings can be viewed but not
changed. The password query appears on leaving a setting table
with the Save and quit command. Resetting the system
configuration to default is also protected.
Calibration password
protects against unauthorized calibration.
If this password is active, the password query appears on calling up
calibration with
c
. If no valid password is entered, calibration is
blocked.
Configuring
passwords
Activating/deactivating and changing passwords is done in the
SETTINGS -> System menu.
Fig. 4-11 SYSTEM SETTINGS menu
Operating notes
The settings password is configured with the Settings
a
/
z
menu
item.
The calibration password is configured with the Calibrate
a
/
z
menu item.
Any activating, deactivating or changing action requires the entry of