Relay outputs
System 182
6 - 2
ba76025e01
02/2012
6.2
Linking and adjusting: general proceedings
6.2.1
Linking relay outputs
Linking options
You have the following options of linking relay outputs:
Sensor S01
The output is linked with sensor 01
Sensor S01
The output is linked with sensor 02
Sensor S01&S02
The output is linked to both sensors S01 and S02.
System
Select this option if you want to use the output for system monitoring
(details, see section 6.4.2).
Proceeding: linking
outputs
Proceed as follows if the output was not yet linked. If you want to
change a link you have to erase the existing link first.
1
Open the SETTINGS menu with
s
.
2
Using
w
and
g
, select and confirm the Relay output Rx
menu item (corresponding to sensor 1, 2 or 3). A list with all
linking options is displayed.
3
Select the required option with
w
and confirm with
g
.
The output is linked. The link is displayed in the header when
the setting of the output is edited.