Encoded data for field bus communication

IQ S

ENSOR

N

ET

Field bus linking

32

ba76148e01

06/2013

5.2

Sensor status

The sensor status applies to both the main and secondary measured
value of a sensor.

Code

Status

Meaning of the status

00h

UNUSED_ID

Sensor and sensor number are not available in the IQ S

ENSOR

N

ET

system.

The measured value status of the main and secondary measured
value is MISSING (4h) (see section 5.6 M

EASURED

VALUE

STATUS

).

The UNUSED_ID status also occurs when the field bus interface of
the IQ S

ENSOR

N

ET

does not receive any data from the

IQ S

ENSOR

N

ET

controller for more than 2 minutes. Possible cause:

Communication malfunction or controller failure. The data transmitted
last remain frozen for the two minutes delay time.

01h

INACTIVE

The sensor is currently inactive. The sensor number (ID) and respec-
tive setting dataset are available in the IQ S

ENSOR

N

ET

system. The

sensor was removed from the IQ S

ENSOR

N

ET

system or the commu-

nication does not work.

The measured value status of the main and secondary measured
value is MISSING (4h) (see section 5.6 M

EASURED

VALUE

STATUS

).

The measured value display on the terminal indicates Init or Error.