1 gsd file for profinet, 2 eds file for ethernet-ip, Configuring the i – YSI IQ S ENSOR N ET Field bus linking User Manual

Page 18: Gsd file for profinet, Eds file for ethernet-ip, Section 3.4, Section

Advertising
background image

Configuring the IQ S

ENSOR

N

ET

IQ S

ENSOR

N

ET

Field bus linking

16

ba76148e01

06/2013

3.4

Configuring the IQ S

ENSOR

N

ET

for Ethernet field

buses

3.4.1 GSD file for Profinet

The GSD file contains all necessary information on the Profinet module
and is required by the configuration program for the Profinet master.

The current GSD file is provided on the Internet under www.WTW.com.

3.4.2 EDS file for Ethernet-IP

The EDS file contains information on the Ethernet-IP module. It is eval-
uated by some configuration programs.

The current EDS file is provided on the Internet under www.WTW.com.

Setting

Selection/Values Explanation

DHCP

Yes

The IQ S

ENSOR

N

ET

is configured as a DHCP

client.
If a DHCP server is in the network, the
IQ S

ENSOR

N

ET

receives all other network

settings from the DHCP server.

No

The IQ S

ENSOR

N

ET

is not configured as a

DHCP client.
All other settings have to be done as required.

IP address

Address

Permanent IP address for the IQ S

ENSOR

N

ET

in

the LAN (if DHCP No).

Subnet mask

Address

Subnet mask (if DHCP No).
The subnet mask depends on the network size
(for small networks: e.g. 255.255.255.0).

DNS server

Address

Entry for field bus not required.
For a connection with the Internet (if DHCP No),
e.g.:

IP address of the DNS server in the network

Entry of IP address or e.g. 127.0.0.1

Standard gateway

Address

Entry for field bus not required.
For a connection with the Internet (if DHCP No),
e.g.:

IP address of the instrument establishing the
access to the Internet

Entry of IP address or e.g. 127.0.0.1

Advertising