Field bus linking

Installing the interface in the IQ S

ENSOR

N

ET

11

ba76148e01

06/2013

Fig. 2-4

Ethernet connection via terminal strip

5

Strip the Ethernet cable for approx. 2 cm and untwist the Rx+,
Rx-, Tx+ and Tx wires.

6

Carefully slash the cable shield (foil + netting) lengthwise and
put it backwards over the cable sheath (pos. 3 in Fig. 2-4).

7

Feed the Ethernet cable through the cable gland into the mod-
ule housing.

8

Where the cable shield was put backwards, press the Ethernet
cable into the shielding terminal (pos. 4 in Fig. 2-4). The shield-
ing terminal must contact the cable shield across a wide area.

9

Connect the Rx+, Rx-, Tx+ and Tx- wires to the LSA terminal
strip with the aid of an LSA punch down tool. Make sure that
the cable assignment agrees with the specification on the ter-
minal label under the terminal strip.

10

Fasten the cap nut of the screwed cable gland.

11

Close the module.

LSA terminal strip

3

4

X10

X9

X8

X7

Rx-

Rx+

Tx-

Tx+

SHIELD CLAMP

ENTHERNET CABLE