2 installing the interface in the iq sensor net, 1 installing the rs485 interface, Installing the interface in the – YSI IQ S ENSOR N ET Field bus linking User Manual

Page 7: Installing the rs485 interface, Section 2.1, 2installing the interface in the iq s

Advertising
background image

IQ S

ENSOR

N

ET

Field bus linking

Installing the interface in the IQ S

ENSOR

N

ET

5

ba76148e01

06/2013

2

Installing the interface in the
IQ S

ENSOR

N

ET

2.1

Installing the RS485 interface

Profibus and Modbus use the same interface (RS485) for data
exchange.

We especially recommend the "Aufbaurichtlinie PROFIBUS/FMS",
published by the PROFIBUS user organization PNO, (PNO order num-
ber 2.111) with many practical tips for installation that also apply to
Modbus systems (

http://www.profibus.com/download/installa-

tion-guide/

).

Further instructions for the installation and operation of a Modbus net-
work are availabe on the Internet under www.modbus.org.

Detailed information on the subject of PROFIBUS is given on the Inter-
net by the PROFIBUS user organization (PNO) under www.profi-
bus.com.

Connecting the Profibus

/ Modbus cable

The Profibus / Modbus cable is connected to the IQ S

ENSOR

N

ET

using

a 9-pin D-SUB connector on the top of the housing.

Suitable cables

Connectors to be used

To remove the blind plug, lift the safety bracket with a suitable screw
driver at both clamps on the left and right side (see Fig. 2-1).

Field bus type

Cable

Profibus

PROFIBUS cable

Modbus

Cable for RS485 (e.g. PROFIBUS cable)

Manufacturer

Type

Phoenix Contact GmbH &
Co. KG
Flachsmarkt 8
32825 Blomberg
Germany
http://www.phoenixcon-
tact.com

VS-09-PROFB-SC
(Phoenix article no. 1654549, with
screwed contacts,
available from YSI under the order no.
902 888Y)

VS-09-PROFB-SP
(Phoenix article no. 1654345, with spring
contacts)