YSI IQ S ENSOR N ET Field bus linking User Manual
Page 31
IQ S
ENSOR
N
ET
Field bus linking
Communication with field buses
29
ba76148e01
06/2013
12
42h
Contents of register 13 (HI)
= main measured value
429E46C2h (Float 32)
--> 79,1382
Measured parameter
and unit,
see byte 10 (measur-
ing mode)
13
9Eh
Contents of register 13 (LO)
= main measured value
14
46h
Contents of register 14 (HI)
= main measured value
15
C2h
Contents of register 14 (LO)
= main measured value
16
00h
Contents of register 15 (HI)
= secondary measured
value
00000000h (Float 32)
--> 0
but measured value
invalid (MISSING),
see Byte 11 (mea-
sured value status)
17
00h
Contents of register 15 (LO)
= secondary measured
value
18
00h
Contents of register 16 (HI)
= secondary measured
value
19
00h
Contents of register 16 (LO)
= secondary measured
value
20
23h
CRC (HI)
Checksum (CRC)
21
5Eh
CRC (LO)
Modbus response
Byte Value
Information
Meaning of the value