Field bus linking

Communication with field buses

29

ba76148e01

06/2013

12

42h

Contents of register 13 (HI)
= main measured value

429E46C2h (Float 32)
--> 79,1382
Measured parameter
and unit,
see byte 10 (measur-
ing mode)

13

9Eh

Contents of register 13 (LO)
= main measured value

14

46h

Contents of register 14 (HI)
= main measured value

15

C2h

Contents of register 14 (LO)
= main measured value

16

00h

Contents of register 15 (HI)
= secondary measured
value

00000000h (Float 32)
--> 0
but measured value
invalid (MISSING),
see Byte 11 (mea-
sured value status)

17

00h

Contents of register 15 (LO)
= secondary measured
value

18

00h

Contents of register 16 (HI)
= secondary measured
value

19

00h

Contents of register 16 (LO)
= secondary measured
value

20

23h

CRC (HI)

Checksum (CRC)

21

5Eh

CRC (LO)

Modbus response

Byte Value

Information

Meaning of the value