2 configuring the iq sensor net for modbus rtu, 1 setting the modbus interface parameters, Configuring the i – YSI IQ S ENSOR N ET Field bus linking User Manual
Page 15: Setting the modbus interface parameters, Section 3.2
IQ S
ENSOR
N
ET
Field bus linking
Configuring the IQ S
ENSOR
N
ET
13
ba76148e01
06/2013
3.2
Configuring the IQ S
ENSOR
N
ET
for Modbus RTU
3.2.1 Setting the Modbus interface parameters
Opening the setting
menu for Modbus RTU
Settings
1
Open the setting menu with <S>.
2
For MIQ/MC2:
Press <> and <OK> to select and confirm the menu
item, Settings of bus interfaces
3
Select and confirm the bus interface with <> and
<OK>.
Setting
Selection/Values
Explanation
Device Address
1 ... 247
Instrument address
of the
IQ S
ENSOR
N
ET
in
the Modbus RTU
Baud rate
1200
2400
4800
9600
19200
38400
57600
Baud rate
Parity
None
Even
Odd
Parity
None
(2 stop bits)
Even
(1 stop bit)
Odd
(1 stop bit)
4
Edit the settings.
To accept all settings, you have to highlight the Save and
quit menu item at the lower end of the setting table and to
confirm with <OK>. If you exit the setting table via <M>,
<ESC>/Quit or the Quit menu item, all changes are
ignored.