YSI IQ SensorNet MIQ/A-MOD Module User Manual
Page 26
Advertising

Data transmission
MIQ/(A-)MOD
26
ba76021e01
01/2012
Determining the number
of registers to be read
The number of registers to be read is maximum 240.
Example: Read all data of X sensors
Example of a
Modbus query
and Modbus response
Query of all sensor information of the sensor with the sensor number
S02:
Number of sensors
(X)
Number of registers
1
8
2
16
3
24
...
...
30
240
Modbus query
Byte
Value
Information
Meaning of the value
1
01h
Modbus address of
the MIQ/MC-MOD
01h --> 1
2
04h
Function
04h --> 4
Read Input Register
(see section 4.3)
3
00h
Start address HI
0008h --> 8
Start with register 9
(Modbus counting method)
4
08h
Start address LO
5
00h
Number of HI registers
0008h --> 8
8 registers
6
08h
Number of LO
registers
7
70h
CRC (HI)
Checksum (CRC)
8
0Eh
CRC (LO)