YSI IQ SensorNet MIQ/A-MOD Module User Manual

Page 26

Advertising
background image

Data transmission

MIQ/(A-)MOD

26

ba76021e01

01/2012

Determining the number

of registers to be read

The number of registers to be read is maximum 240.

Example: Read all data of X sensors

Example of a

Modbus query

and Modbus response

Query of all sensor information of the sensor with the sensor number
S02:

Number of sensors
(X)

Number of registers

1

8

2

16

3

24

...

...

30

240

Modbus query

Byte

Value

Information

Meaning of the value

1

01h

Modbus address of
the MIQ/MC-MOD

01h --> 1

2

04h

Function

04h --> 4
Read Input Register
(see section 4.3)

3

00h

Start address HI

0008h --> 8
Start with register 9
(Modbus counting method)

4

08h

Start address LO

5

00h

Number of HI registers

0008h --> 8
8 registers

6

08h

Number of LO
registers

7

70h

CRC (HI)

Checksum (CRC)

8

0Eh

CRC (LO)