Installation

MIQ/Blue PS

3 - 20

ba76095e02

01/2013

3.6.2

Connection to 24 V DC

Warning

If the 24 V DC supply is connected incorrectly, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The MIQ/Blue PS may only be connected by a trained electri-

cian.

The 24 V DC supply must meet the specifications quoted on the

nameplate and in chapter 6 T

ECHNICAL

DATA

(protective low volt-

age SELV).

The connection of the MIQ/Blue PS to the power supply may

only be carried out when it is not carrying any voltage.

When installed in a building, a switch or power switch must be

provided as an interrupt facility for the MIQ/Blue PS.
The interrupt facility must

– be installed in the vicinity of the MIQ/Blue PS, easily acces-

sible by the user, and

– be labeled as an interrupt facility for the MIQ/Blue PS.

Note
Battery systems should be protected against total discharge. The MIQ/
Blue PS has no integrated deep discharge protection.

Materials required

 Wire end sleeves, suitable for the 24 V AC/DC feed line, with suit-

able crimping tool

 1 x cable gland, suitable for the cable diameter (see section 3.5 on

page 3-14).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screw driver

 Small screw driver.

Preparing the 24 V DC

line

1

Cut off the cable to the required length.

2

Strip the cable insulation for approx. 45 mm.

3

Bare the wires 1 and 2 and fit them with wire end sleeves.