TriOxmatic

®

702 IQ

Commissioning

3 - 1

ba76048e01

01/2012

3

Commissioning

3.1

Scope of delivery

 TriOxmatic

®

702 IQ

 ZBK 25i accessory case

 RA 600 cleaning attachment

 The sensor is filled with electrolyte and provided with a protective

hood.

 Operating manual.

3.2

Installation

Connection cable

The SACIQ sensor connection cable is required to connect the sensor.
Information on this and other IQ S

ENSOR

N

ET

accessories is given in

the YSI catalog and on the Internet.

Note
How to connect the SACIQ sensor connection cable to the terminal
strip of an MIQ module is described in chapter 3 I

NSTALLATION

of the

IQ S

ENSOR

N

ET

system operating manual.

Are the plug

connections dry?

Before connecting the sensor and sensor connection cable, please
make sure the plug connections are dry. If moisture gets into the plug
connections, first dry the plug connections (dab them dry or blow them
dry using compressed air).

Note
Do not suspend the sensor on the sensor connection cable. Use an
armature or flow-thru vessel. We recommend to perform the
measurement and installation in the D 702/N flow-thru vessel using the
ADA-DF 4 and ADA-DF 6 adapters.
Information on this and on other IQ S

ENSOR

N

ET

accessories is given

in the YSI catalog and on the Internet.

Minimum approach flow

The minimum required approach flow at the sensor must be present
during measurement (see chapter 7 T

ECHNICAL

DATA

). The minimum

approach flow can, e.g. be achieved by:

 the flow rate of the water

 the correct flow-thru in the flow-thru vessel.