Technical data

7 - 1

ba76048e01

01/2012

7

Technical data

7.1

Measurement characteristics

Measuring principle

Membrane-covered amperometric sensor with potentiostatically
operated 3-electrode system;
Integrated microprocessor electronics, shielded 2-wire connection for
power and data transmission.

Electrolyte

ELY/A

Measuring ranges and

resolutions

Temperature

compensation

IMT (integrated membrane temperature compensation)

Measurement in water

According to solubility function DIN 408 / Part 22

Measurement in salt-

contaminated

wastewater

Salinity input from 2.0 ... 70.0;
corresponds to 3.4 mS/cm ... 86.2 mS/cm at T

REF

20 °C

(salinity measurement in the range 2 ... 42 according to IOT =
International Oceanographic Tables)

Polarization time

Temperature

measurement

Calibration procedure

Air calibration

Measuring
mode

Measuring range

Resolution

D. O.
concentration

0 ... 2000 µg/l
0.00 ... 10.00 mg/l

1 µg/l
0.01 mg/l

D. O. saturation

0.0 ... 110.0 %

0.1 %

In the case of recommissioning or
change of electrolyte

At least 60 minutes

for precision measurements

At least 120 minutes and
recalibrate on the following day

In the case of short polarization
interruptions (depending on the
duration of the interruption)

30 to 120 minutes

Temperature sensor

integrated NTC

Measuring range

- 5 °C ... + 60 °C (23 ... 140 °F)

Accuracy

± 0.5 K

Resolution

0.1 K