2 cleaning the sensor shaft and membrane, Cleaning the sensor shaft and membrane -2, Ne (see section 5.2 c – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 24: Leaning, Sensor, Shaft, Membrane, Section 5.2) and

Advertising
background image

Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

702 IQ

5 - 2

ba76048e01

01/2012

5.2

Cleaning the sensor shaft and membrane

For normal operation (e.g. boiler feed water), cleaning and calibration
are strongly recommended:

 if the measured values are suspected of being incorrect (usually too

low)

 if the measured value lies outside the range of precision required by

the user during the function check.

Cleaning agents

Cleaning

Caution

Acetic acid irritates the eyes and the skin. When handling acetic
acid, always wear protective gloves and protective goggles.

Caution

Clean the membrane very carefully!
Make sure that the membrane is not damaged, e.g. by sharp-
edged objects.

Contamination

Cleaning agents

For loosely adhering dirt or
biological deposits

Soft cloth or soft sponge, warm tap
water with detergent

Salt and / or lime deposits

Acetic acid (volume percentage =
20 %),
soft cloth or soft sponge

1

Take the sensor out of the flow-thru vessel and put it into a
holding device.

2

Carefully clean the sensor shaft and the membrane as
specified in the C

LEANING

AGENTS

section.

3

Then, rinse it thoroughly with tap water.

Advertising