YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual
Page 13
TriOxmatic
®
702 IQ
Commissioning
3 - 3
ba76048e01
01/2012
3.3
Commissioning / Getting the instrument ready for
measuring
Note
For the sensor to polarize, the following conditions must be fulfilled:
The sensor is connected with the IQ S
ENSOR
N
ET
system via the
SACIQ sensor connection cable.
The IQ S
ENSOR
N
ET
system is in operation
The sensor has been identified by the IQ S
ENSOR
N
ET
system.
Precision
measurements
Recommendation: In order to carry out precision measurements, leave
the sensor to polarize for a longer period of time, e.g. overnight, and
recalibrate it the following day.
1
Pull the protective cap off the sensor.
2
Leave the sensor to lie in the air for at least 60 minutes while it
is switched on (polarization).
3
If required, assign a user-defined name to the sensor (see
relevant IQ S
ENSOR
N
ET
system operating manual).
4
Set the sensor (see section 3.4).
5
Calibrate the sensor (see section 4.2 C