Measuring / Operation

4 - 3

ba76048e01

01/2012

Note
Up to this point, you can break off the calibration procedure at any time
with the

e

key. The system continues to work with the old calibration

data. However, you have to switch off the maintenance condition in any
case.
After starting the calibration with the

g

key (step 4), you cannot break

off the calibration any more.

Note
In the case of air temperatures under 5 °C do not calibrate the sensor
in air but in air-saturated water that has a higher temperature. You
obtain air-saturated water by pouring water several times in and out of
two vessels so that it sparkles. Then immerse the sensor in a vessel
with air-saturated water and calibrate as described above.

Possible results of the

calibration

The calibration data are evaluated by the system. A calibration can
have the following results:

3

Bring the sensor into the calibrating position.
To do so, position the sensor approx. 2 cm above a water
surface, best in a narrow bucket or similar container with water.
When doing so, no liquid should be brought on the membrane.

4

Continue with

g

.

The calibration data is automatically determined and
documented in the calibration record.

5

Complete the calibration procedure according to the
instructions on the display.

Display after the
calibration

Log book entries
(meaning/actions)

Measured value display

Sensor was successfully calibrated.
For the calibration data, see the calibration
history.

"----"

Sensor could not be calibrated.
Sensor blocked for measurement.
– Service the sensor immediately

(see operating manual).

– View calibration history.

– Check calibration conditions and

calibration standard.