6 what to do if, What to do if -1, 6what to do if – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 41

Advertising
background image

TriOxmatic

®

702 IQ

What to do if...

6 - 1

ba76048e01

01/2012

6

What to do if...

The sensor is in the air

and the display shows

0.0 mg/l or 0 % O

2

The sensor cannot be

calibrated

If the sensor still cannot

be calibrated after

changing the electrolyte

and membrane cap

Mechanical damage to

the sensor

Measured values too low

Cause

Remedy

– No electrolyte in the membrane

head

– Change the WP 25i

membrane head (see
section 5.3)

Cause

Remedy

– Contaminated membrane head

– Clean the outside of the

sensor (according to
operating manual), wait at
least 15 minutes and
recalibrate it.

– In the case of contamination

that cannot be removed:
Change the membrane
head and electrolyte

Cause

Remedy

– Contaminated electrodes or

sensor toxification

– Clean the electrodes (see

section 5.4)

Cause

Remedy

– Return the sensor

Cause

Remedy

– Membrane contaminated

– Sensor has not been calibrated

for a long time

– Clean the outside of the

sensor, then recalibrate (see
section 5.2 and section 4.2)

– Membrane does not fit snugly on

the gold working electrode

– Change the membrane cap,

then recalibrate (see section
5.3 and
section 4.2)

– Minimum approach flow not

present

– Ensure sufficient approach

flow

Advertising