2 cleaning the silver counter electrode, Cleaning the silver counter electrode -10, E (see section 5.4.2) – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 32

Advertising
background image

Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

702 IQ

5 - 10

ba76048e01

01/2012

5.4.2

Cleaning the silver counter electrode

Caution

The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective. However, no danger can arise if the RA 600 cleaning
attachment is used correctly.

Cleaning the silver

counter electrode with

the RA 600

cleaning attachment

Caution

Never grease or lubricate the inner O-ring in the RA 600 cleaning
attachment!

1

Wipe the silver counter electrode with a lint-free paper towel
and carefully remove any loose deposits.

2

Screw the cleaning attachment onto the sensor, instead of the
membrane head.