Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
702 IQ
5 - 6
ba76048e01
01/2012
Caution
The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.
7
Throw away the first filling and fill the membrane head once
more with electrolyte solution.
8
Remove any air bubbles by carefully tapping the membrane
head.
9
Screw the membrane head onto the shaft while holding the
sensor at an angle. Excess electrolyte solution is forced out of
the ventilation area.
Ventilation area (see above)