Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

702 IQ

5 - 6

ba76048e01

01/2012

Caution

The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.

7

Throw away the first filling and fill the membrane head once
more with electrolyte solution.

8

Remove any air bubbles by carefully tapping the membrane
head.

9

Screw the membrane head onto the shaft while holding the
sensor at an angle. Excess electrolyte solution is forced out of
the ventilation area.