TriOxmatic

®

702 IQ

Maintenance, cleaning, disposal, and replacement

5 - 1

ba76048e01

01/2012

5

Maintenance, cleaning, disposal, and
replacement

5.1

General maintenance instructions

Warning

Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective measures
must be taken (protective clothing, protective goggles, etc.).

Maintenance condition

We recommend to switch on the maintenance condition each time
before removing the sensor from its measuring position. This avoids
unintended reactions of linked outputs. For more detailed information
on the maintenance condition please refer to the respective
IQ S

ENSOR

N

ET

system operating manual.

When to unscrew the

sensor from the

connection cable?

To clean the outside of the sensor (section 5.2) and to replace the
electrolyte solution and membrane head (section 5.3), we recommend
to leave the sensor connected to the sensor connection cable.
Otherwise, moisture and/or dirt can get into the plug connection where
they can cause contact problems. If you wish to disconnect the sensor
from the sensor connection cable, note the points described below.

For the cleaning of the electrodes (section 5.4) the sensor must be
unscrewed from the sensor connection cable to prevent any damage.
Observe the following points for this:

 Unscrew the sensor from the SACIQ sensor connection cable.
 Always place a protective cap on the sensor plug head connector

and on the SACIQ sensor connection cable so that no moisture or
dirt can get into the contacting surfaces.

 In corrosive environments close the dry socket of the sensor

connection cable with the SACIQ-Plug protective screw cap in order
to protect the electrical contacts from corrosion. The protective cap
is available as an accessory (see section 5.8 M

AINTENANCE

EQUIPMENT

AND

REPLACEMENT

PARTS

).

 Connect the sensor with the SACIQ sensor connection cable again

before polarization.