6 what to do if, What to do if -1, 6what to do if – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 45

Advertising
background image

TriOxmatic

®

701 IQ

What to do if...

6 - 1

ba76047e01

01/2012

6

What to do if...

The sensor is in the air

and the display shows

0.0 mg/l or 0 % O

2

The sensor cannot be

calibrated

If the sensor still cannot

be calibrated after

changing the electrolyte

and membrane cap

Mechanical damage to

the sensor

Measured values too low

Measured value

fluctuating heavily

Cause

Remedy

– No electrolyte in the membrane

head

– Change the WP 25i

membrane head (see
section 5.3)

Cause

Remedy

– Contaminated membrane head

– Clean the outside of the

sensor (according to
operating manual), wait at
least 15 minutes and
recalibrate it.

– In the case of contamination

that cannot be removed:
Change the membrane
head and electrolyte

Cause

Remedy

– Contaminated electrodes or

sensor toxification

– Clean the electrodes (see

section 5.4)

Cause

Remedy

– Return the sensor

Cause

Remedy

– Membrane contaminated

– Sensor has not been calibrated

for a long time

– Clean the outside of the

sensor, then recalibrate (see
section 5.2 and section 4.2)

– Membrane does not fit snugly on

the gold working electrode

– Change the membrane cap,

then recalibrate (see section
5.3 and
section 4.2)

Cause

Remedy

– Membrane cap loose

– Screw the membrane head

tight

– Membrane does not fit snugly on

the gold working electrode

– Change the membrane cap,

then recalibrate (see section
5.3 and
section 4.2)